FC
Parameter(s): Free chlorine, chlorine dioxide, and/or dissolved ozone
OVERVIEW
FREE CHLORINE ANALYZER
Real Tech’s bypass style Free Chlorine analyzer provides rapid real-time measurement of free chlorine, chlorine dioxide and/or dissolved ozone in water or wastewater. Ideal for pressurized applications, the monitoring solution has three measurement ranges.
FEATURES & BENEFITS
- Affordable, real-time free chlorine, chlorine dioxide and/or dissolved ozone measurement
- Potentiostatic sensor technology
- Range for wastewater to high purity applications
- Automatic temperature compensation
- Low maintenance requirements
- Simple to install and operate
- No reagents required
SPECIFICATIONS
CHARACTERISTIC | TECHNICAL DATA |
SENSOR | |
Parameters | Free chlorine, chlorine dioxide, and/or dissolved ozone |
Free Chlorine Range | Selection of three measuring ranges: 0-2 mg/L , 0-20 mg/L or 0-200 mg/L |
Resolution | 0.001, 0.01 or 0.1 |
Units | ppm / mg/L |
Sampling Time | 0.5 seconds |
Measurement Principle | Potentiostatic measurement technology |
Electrodes | 2 platinum rings |
Reference | Gel with annular junction |
Sensor Lifetime | Approximately 12-18 months. Depends on site and application |
Calibration | Manual correction on the selected scale and sensor’s current offset |
Cleaning | Pressurized DI water only |
Temperature Compensation | Automatic with RTD |
Flow | Constant flow required. ~150-500 mL/min |
Fluid Connections | Inlet: 1/4” fitting. Outlet: fitting for 10x14 mm tubing |
Dimensions | 5.90”H x 3.54”W x 1.57”D |
TRANSMITTER | |
Interface | Four buttons to control set-up and configuration |
Display | 1 line x 8 character LCD |
Output | Self-powered 4-20mA, RS485 serial to PC |
Electrical | 9-36 VDC |
Operating Temp | 0 to 50°C (32 to 122°F) |
Storage Temp | -10°C to 60°C (14 to 140°F) |
Weight | 5 lb |
Enclosure | IP40, wall mountable. IP65 / NEMA4 optional. |
Dimensions | 2.79”H x 3.74”W x 2.28”D (IP65/ NEMA 4 cabinet - 10.25”H x 10.25”W x 5.5”D) |
Warranty | 2-year limited warranty |
* Technical Specifications are subject to change without notice.
Note: The free chlorine sensor requires a constant flow for stable measurement, therefore an accompanying bypass manifold (chamber) is included. The bypass manifold and transmitter are both wall mountable.
MODEL & RANGE
FC model includes potentiostatic sensor, temperature sensor, transmitter and bypass chamber.
MODEL# | FREE CHLORINE (mg/L) |
FC | Three measuring ranges: 0-2 mg/L , 0-20 mg/L or 0-200 mg/L |
* Stated ranges approximate, dependent on industry, site and application
ACCESSORIES
ACCESSORIES
PRODUCT# | NAME | DESCRIPTION |
S-211001 | FC IP65 Enclosure | IP65 NEMA 4 enclosure upgrade for free chlorine sensor transmitter |